The accuracy of hundreds of MRI scans conducted at a major Melbourne hospital have been called into question in a major health scandal.

Melbourne’s Maroondah Hospital was forced to launch an investigation after it was discovered an MRI machine had been producing compromised images.

The issue is understood to have been occurring since January 2025, with insiders telling the Herald Sun that between 900 and 1000 scans may have been impacted.

An MRI machine is used for patients with a wide range of conditions, with those affected including patients needing breast screenings, as well as abdomen and spine scans.

Some Maroondah Hospital patients have already been forced to undergo repeat testing at other hospitals, and insiders have said patients “should request to review all their scans for mispathology”.

The problem in Maroondah is being caused by the failure of the Faraday cage built around the MRI machine, which is designed to shield it from electromagnetic interference.

When an investigation was held to assess the issue, Maroondah Hospital’s MRI machine and its Faraday cage failed on 69 of the 72 performance measures, with the findings showing scans risked being compromised.

The issue has led staff at the hospital to resort to using aluminium foil and cardboard boxes to block outside equipment interfering with quality of images produced by the MRI machine.

However Eastern Health executive director Toni Gutschlag stressed that patient safety has not been compromised by the issue and only a small number of scans had been impacted.

“Eastern Health takes the safety and quality of our diagnostic services extremely seriously, and patient care remains our highest priority,” she said.

“While an RF shielding (Faraday cage) issue has been identified, only a small number of scans are impacted, and all images are rigorously reviewed to ensure diagnostic accuracy, patient safety has not been compromised.”

Liberal MP Nick McGowan, whose upper house seat includes Maroondah Hospital, said it was unacceptable that the issue had not been addressed.

“Something at Eastern Health is going drastically wrong when they are allowing lesser quality imaging for months on end,” he said.

Mr McGowan called on the Allan government to immediately replace both the Faraday cage and the MRI machine, pointing out the government had known about the issue for at least six months.

“It is completely unacceptable that many local women, that anybody, is subjected to substandard MRI facilities because the state government refuses to fund a replacement shield and provide the world’s most advanced MRI machine,” he said.

“By ignoring this leaking cage at Maroondah Hospital, the state government is effectively telling eastern suburbs families, my families, that their diagnostic precision isn’t worth the budget expense.”

Health Minister Harriet Shing said the Allan government had provided adequate funding to the hospital and it was the hospital’s responsibility to monitor and maintain the condition of equipment.

“Health services are required to monitor and maintain the condition of all equipment including equipment used in diagnostic services – it’s as simple as that,” she said.

“Eastern Health has advised the MRI machine is still operational, and patient care has not been impacted.

“Our health services continue to receive record funding – ensuring every public hospital has what they need to keep delivering world-class care to Victorians.”

But shadow health minister Georgie Crozier said response from the government was “pathetic”.

The Liberal MP, who previously worked in the health sector, said having substandard equipment at a major Melbourne hospital was “very concerning”.

“It’s not only the potential for a misdiagnosis, or not picking up something that’s in a scan,” she said.

“What about the duplication, the repeat scans that have got them done? What about the cost of that? And what about the anxiety to the patient that that causes?

“This is substandard, it’s not good enough, and I think the minister’s excuse is pathetic.”