Maxim Berg/Unsplash.com
CGI said Tuesday that CGI AI Head Analysis, software that flags suspected brain bleeds on head CT scans, earned EU MDR class IIb certification, but the company released no sensitivity or specificity figures for the certified version. The best public numbers come from a 2025 Scientific Reports study by the same Helsinki University Hospital (HUS) team, and they imply that about one flag in nine pointed to a real bleed, a false-alarm load that a hospital deploying a tool with those numbers would absorb.
CGI developed the software with HUS and Planmeca, a Helsinki-based maker of imaging equipment, and issued the announcement at 6:30 a.m. ET from Helsinki. The tool reads head CTs for non-traumatic, spontaneous intracranial hemorrhage and marks suspected acute blood so a physician can review it. CGI pitches it for busy emergency departments, where a doctor wants to rule out a missed and potentially fatal bleed.
The legal status is new. In May 2023, CGI said the tool, then called CGI AI Head, was not yet CE-marked and was not yet available. Finland alone performs about 180,000 head CTs a year.
Certification Arrived With No Sensitivity or Specificity FiguresProductCGI AI Head AnalysisRegulationEU Medical Device Regulation (MDR), class IIbFunctionFlags suspected acute, spontaneous (non-traumatic) intracranial bleeding on head CT for physician reviewDevelopersCGI, Helsinki University Hospital (HUS), PlanmecaNotified bodyNot named in the releasePerformance dataNone given: no sensitivity, specificity, dataset size or validation detailsLaunch date, pricing, customersNot given
The release says certification lets hospitals deploy the tool at scale and confirms it meets strict national and EU-level standards for safety, quality and performance. It says scientific articles about the solution appeared in internationally respected journals but names none. CGI’s case-study page cites Neurology and Nature, while the Helsinki group’s 2025 paper appeared in Scientific Reports.
CGI’s own materials add that the covered bleed types account for more than 99% of acute brain bleeds and that the software identifies a specific bleed with up to 98% accuracy. The page does not define accuracy or name a dataset, so those figures cannot be compared with the studies below. The tool targets non-traumatic bleeds, and trauma-related hemorrhage is outside its stated use.
Does EU MDR Certification Prove Brain Bleed AI Works?
Not by itself. Under MDR Rule 11 software classification, software that supplies information for diagnostic decisions starts at class IIa, rises to IIb when those decisions could cause serious deterioration of health or a surgical intervention, and reaches class III if they could cause death or irreversible deterioration. Higher-risk devices, including class IIb, need a notified body, an organization designated by an EU member state, to assess conformity before the manufacturer can apply the CE mark. CGI did not say which body certified the tool.
A CE mark affirms conformity with regulatory requirements; it is not a quality indicator, and it does not show clinical benefit. Critics of the notified-body system have long cited uneven expertise and forum shopping for lenient bodies, concerns that predate the MDR. CGI has not published the technical file behind this decision, and its release does not mention the U.S. Food and Drug Administration (FDA); a CE mark is a European, not an American, authorization.
CGI is not the first head CT bleed detector with an MDR class IIb certificate. Annalise.ai announced class IIb for its head CT product in August 2023, and Qure.ai announced class IIb for a portfolio that includes head CT in February 2023.
Four U-Nets, One Metamodel and Pixel Threshold Decide Each Flag
The published Helsinki system works as follows, and CGI has not said whether the certified build matches it. The 2025 paper describes an ensemble of four U-Net networks whose outputs feed a metamodel, followed by rule-based post-processing, trained on about 300 head CTs in total. A U-Net, introduced in 2015 by Olaf Ronneberger, Philipp Fischer and Thomas Brox, is a convolutional network built for image segmentation, which outlines regions within a scan instead of only labeling the whole scan.
A fixed rule then converts pixels into an alert. Post-processing discards clusters smaller than 10 pixels, and a scan counts as positive when the combined cluster exceeds 125 pixels. That threshold sets the trade-off: sensitivity and specificity generally move in opposite directions, so catching fainter bleeds means accepting more false flags. The Helsinki group’s earlier subarachnoid-only model shows the effect: on 1,379 external cases it found 136 of 137 bleeds, a 99.3% sensitivity, at 63.2% specificity.
CGI’s case study says the radiologist first analyzes the images independently and then compares the result with the AI output. CGI said in 2023 that it built the platform on Microsoft Azure. The certified version’s alert threshold and bleed-type coverage have not been published, so hospitals cannot yet see where the dial sits that sets their false-alarm load.
Helsinki Study Caught Every Bleed Within 12 Hours and 77.8% at 12 to 24 Hours
The Scientific Reports paper tested the ensemble on 7,797 consecutive non-contrast emergency head CTs from 10 emergency rooms in the HUS area, imaged from Oct. 1 to Dec. 31, 2021. On-call radiologists had reported 118 of them as showing spontaneous hemorrhage. Overall, the software reached 89.8% sensitivity (95% confidence interval 84.4% to 95.3%) and 89.5% specificity (88.8% to 90.2%). It detected all 78 hemorrhages imaged within 12 hours of symptom onset, 19 of them with timing presumed, reached 77.8% sensitivity in 18 patients imaged 12 to 24 hours after onset, and flagged five hemorrhages that on-call radiologists had missed.
The dip at 12 to 24 hours may not be only the software. CT itself loses sensitivity for subarachnoid hemorrhage as hours pass, identifying about 98.7% of cases within six hours of onset.
An earlier HUS algorithm for subarachnoid hemorrhage alone, described in Neurology in 2023, trained on 90 positive scans and 22 negatives. On 519 consecutive emergency scans from five Finnish hospitals, it found all 8 subarachnoid bleeds, with 75.3% specificity reported in the abstract, although the paper’s results text gives 0.87. Its authors reported no relevant disclosures. For the 2025 paper, the authors declare no competing interests, although one author is affiliated with Planmeca, a co-developer of the product. CGI’s release does not say whether the certified software matches either published algorithm.
Low Prevalence Turns 89.5% Specificity Into About 1 in 9 Correct Flags
Hemorrhage is rare among emergency head CTs: 118 of 7,797 scans, about 1.5%. A specificity of 89.5% across the 7,679 scans radiologists reported as negative implies roughly 800 flags on those scans. Against about 106 true detections, that works out to about one correct flag in nine. This is a calculation from the study’s headline figures, not a number the authors report, and it may slightly understate precision if the five bleeds radiologists missed were counted as false flags.
In plain counts, a hospital with the same case mix would see about 15 bleeds per 1,000 scans, about 14 of them flagged, along with about 103 false flags. False alarms are not necessarily a problem for a triage tool, since a flag sends a scan to a physician who still decides. But a high false-positive rate can wear down trust and add reading time, and the burden depends on the threshold a deployed system uses, which CGI has not disclosed.
Prevalence drives that arithmetic. Sensitivity and specificity are prevalence-independent properties of a test, while positive predictive value changes with how common the condition is. When a condition is rare, the false positive paradox can leave false alarms outnumbering true ones. A Danish emergency-department study of 844 scans with 7.5% bleed prevalence found that RAPID ICH, a commercial tool, reached 87.3% sensitivity, 74% specificity and 21.3% positive predictive value, with 203 false positives most often tied to dural thickening and choroid plexus calcification. The authors wrote that the low predictive value limited generalizability and raised workload and alert-fatigue concerns.
Radiologists describe the same burden. In a survey of 65 radiologists using Avicenna.ai’s CINA, an FDA-cleared tool, at a national teleradiology practice, 33.8% agreed that time spent reviewing false-positive alerts outweighed the benefits, including 52.2% of neuroradiologists, and only 10.8% reported shorter interpretation times. The survey measured perceptions at one practice, not diagnostic behavior.
Published specificity for other tools sits higher, with wide variation. A systematic review of studies through May 2025 reported 89.9% sensitivity and 95.1% specificity for commercial systems. An external validation of an FDA-approved model on 5,600 scans reported 89% sensitivity and 96% specificity. A 2023 Frontiers in Neurology evaluation of VeriScout on 527 scans in Sydney reported 92% sensitivity and 96% specificity. Those datasets differ, so none is a head-to-head comparison, but specificity in the mid-90s is the yardstick the HUS tool’s 89.5% will be read against.
Hospitals Should Ask CGI for Alert Threshold, Local Test Results and Outcome Data
Before deployment, hospitals can ask CGI for the certified version’s sensitivity and specificity, its alert threshold, the expected false flags per 1,000 scans at the hospital’s own bleed prevalence, the notified body and the certified intended use. A 2025 Nature Medicine reporting guideline, STARD-AI, sets out what diagnostic-accuracy studies of AI should report and gives buyers a checklist for those requests.
CGI says the radiologist and the AI first analyze images independently, and it said in 2023 the tool was approved for retrospective testing in HUS’s clinical workflow. That design is meant to limit automation bias, the tendency to favor automated suggestions and disregard conflicting information. In the Avicenna survey, 6.2% (4 of 65) reduced scrutiny after an AI-negative result, none of the neuroradiologists and 9.5% of the other radiologists.
Whether flags speed care is unsettled. In one study published by RSNA, an intracranial hemorrhage flag and widget alone had no significant effect on exam wait time, while actively reprioritizing the reading worklist cut waits for flagged exams. CGI’s release cites no outcome study of the tool’s effect on time to diagnosis or treatment.
Korja Calls Certification Historic While CGI Says Tool Is Ready for Clinical Use
Miikka Korja, a chief physician and associate professor of neurosurgery at HUS who led the project, called the certification a historic achievement and said the algorithms’ scientific basis is exceptionally strong, making the solution highly reliable. Sonja Pietilä, a vice president and consulting expert at CGI who leads solution development, said certification shows the tool is ready for real-world clinical use and can now be built into hospital imaging workflows and decision-making.
HUS ran the project inside CleverHealth Network, an ecosystem it coordinates, established in 2017 to develop health products for the global market, and Korja has said revenue from international-market services could fund specialized care without costing municipalities. In January 2023, HUS openly distributed its subarachnoid algorithm to researchers, put a website online where anyone could test it with a head CT, and said it hoped to test the package clinically in 2023 and later win official approval for patient care.
For CGI, a consulting and IT services firm with 94,000 professionals and fiscal 2026 revenue of CA$16.57 billion (approximately $11.54 billion), the CE mark turns a research collaboration into a product it can sell to hospitals. HUS keeps a digital archive of more than 16 million head scans, according to a regional council description of the project, though the published studies trained on far smaller sets.
From August 2, 2028, EU AI Act Rules Will Apply on Top of MDR to Imaging AI
AI built into regulated products, including medical devices, will face the EU AI Act’s high-risk obligations from August 2, 2028, after the Digital Omnibus on AI moved the date from August 2, 2027. The Council of the EU gave its final approval on June 29, 2026. AI-enabled medical devices remain subject to both the AI Act and the MDR, and MedTech Europe says the deal leaves open how the two regimes fit together.
Certification Licenses Sales While Buyers Still Need Local False-Alarm Numbers
CGI AI Head Analysis can now be sold to EU hospitals as a class IIb medical device. What CGI has not shown is how often its flags are right at a given hospital. The Helsinki team’s best public study caught most bleeds early but implies roughly eight false flags per true detection at a 1.5% bleed rate, and a hospital with a different case mix, scanner or threshold will see a different ratio. Radiologists and emergency physicians should expect to see those numbers before the first alert fires, patients should understand that a flag prompts a physician’s read rather than delivering a diagnosis, and any claim of faster care needs outcome data that has not yet been published.
Currency conversions are approximate and use the exchange rate as of Nov. 24, 2026.
Frequently Asked QuestionsWhat does EU MDR class IIb certification mean for brain bleed AI?
It means a notified body assessed the manufacturer’s conformity under the EU Medical Device Regulation, so the manufacturer can apply a CE mark and sell the software in the EU. It is not a published accuracy result or proof of clinical benefit.
How often are brain bleed AI alerts wrong?
It depends on the tool, its threshold and how common bleeds are among scanned patients. In the Helsinki study, about 1 in 9 flags matched a radiologist-reported bleed at a 1.5% bleed rate, while a Danish study of another tool found 21.3% of alerts correct at a 7.5% rate. Ask any vendor for expected false flags per 1,000 scans at your own case mix.
Does CGI AI Head Analysis replace radiologists?
No. CGI describes the tool as flagging suspected bleeds for physician review, and its materials describe a workflow in which the radiologist reads the images first and compares afterward with the AI results. HUS said in 2023 that diagnoses remain with radiologists and treatment decisions with the attending physician.
Is CGI AI Head Analysis cleared in the United States?
CGI’s announcement covers EU MDR certification only and does not announce any FDA clearance. A CE mark is a European, not an American, authorization.